The Beer-Lambert law
A = ε × l × c
| Symbol | Quantity | Usual unit |
|---|---|---|
| A | Absorbance | none |
| ε | Molar absorptivity (extinction coefficient) | L·mol⁻¹·cm⁻¹ |
| l | Path length of the cuvette | cm |
| c | Concentration | mol/L |
Absorbance is proportional to how much absorbing substance the light passes through: double the concentration or the path length and the absorbance doubles.
02
Worked example
A solution of NADH reads A = 0.622 at 340 nm in a 1 cm cuvette. With ε = 6,220 L·mol⁻¹·cm⁻¹:
c = A ÷ (ε × l) = 0.622 ÷ (6,220 × 1) = 1.00 × 10⁻⁴ mol/L = 100 µM
The transmittance is 10^(−0.622) = 23.9%: about a quarter of the light gets through.
03
Absorbance and transmittance
A = −log₁₀(T) · %T = 100 × 10^(−A)
| Absorbance | Light transmitted |
|---|---|
| 0 | 100% |
| 0.3 | 50.1% |
| 1 | 10% |
| 1.5 | 3.16% |
| 2 | 1% |
04
Where the law stops working
- High absorbance. Above roughly 1 to 1.5 so little light reaches the detector that stray light flattens the response. Dilute and measure again; the dilution calculator helps.
- Concentrated solutions. Molecules interact and ε is no longer constant.
- Scattering. Turbid samples and suspensions scatter light, which reads as absorbance but is not.
- The wrong ε. ε depends on wavelength, solvent and pH. Use the value for your exact conditions.
- For unknowns, a standard curve fitted with the linear regression calculator is more reliable than a literature ε.
05
Path length
Standard cuvettes have a 1 cm path. Microvolume spectrophotometers use about 1 mm or less and usually report absorbance already scaled to 1 cm; plate readers have a path that depends on the volume in the well. Enter the path your absorbance value actually refers to.
06
Common mistakes with the Beer-Lambert law
- Entering the path length in millimetres. A 10 mm cuvette typed as 10 cm gives 10 µM for the NADH example instead of 100 µM, ten times too low.
- Converting transmittance with the natural log. A = −log₁₀(T), so 23.9% T is A = 0.622; −ln(0.239) gives 1.431 and a concentration of 230 µM instead of 100 µM.
- Forgetting to subtract the blank. If the buffer alone reads 0.050 and the sample reads 0.672, using 0.672 gives 108 µM, 8% too high; 0.672 − 0.050 = 0.622 gives 100 µM.
- Trusting very high readings. At A = 1.5 only 3.16% of the light reaches the detector, and at A = 2 only 1%, so small amounts of stray light distort the result. Dilute and measure again.
Trust
How this page is checked
- The calculation runs on code with automated tests against worked reference values. Every number in the examples on this page was produced by that same code, not typed by hand.
- The formulas and conventions used are stated on the page, so you can reproduce the result yourself or compare it with your own software.
- Last reviewed by the Sciloom team, Zebra Nine Studio Inc. It is a tool for research and teaching, not a substitute for a statistician, a validated lab method or medical advice.
Cite this page
Sciloom. (2026). Beer-Lambert law calculator. Zebra Nine Studio Inc. https://sciloom.si/beer-lambert-law-calculator
Frequently asked questions
How do I calculate concentration from absorbance?
Divide the absorbance by the molar absorptivity times the path length: c = A ÷ (ε × l). Leave the concentration box empty and fill in the other three.
What are the units of molar absorptivity?
Litres per mole per centimetre (L·mol⁻¹·cm⁻¹, also written M⁻¹cm⁻¹), so that multiplying by cm and mol/L gives a number without units.
How do I convert absorbance to transmittance?
Percent transmittance is 100 × 10^(−A). An absorbance of 1 means 10% of the light passes through.
Why is my absorbance not proportional to concentration?
Usually because the reading is too high, the sample scatters light, or the solution is concentrated enough for the molecules to interact. Dilute into the range below about 1.