Standard curve calculator

This standard curve calculator fits a straight line to your standards and reads the concentration of unknown samples off it. Paste the standard concentrations, their signals and the signals of your unknowns to get the equation, R² and each unknown’s concentration, with a warning when a sample falls outside the standards.

Your numbers are calculated in your browser and are not sent or stored.

Standard curve calculatorOr describe it in Studio →

Subtract the blank from every reading first, or include the blank as a zero-concentration standard, not both.

Standard curve

signal = 0.03425 + 0.00097 × conc

R² = 0.9996

#SignalConcentration
10.4377.1 µg/mL
20.85841 µg/mL
01000 µg/mL

To report

Standard curve: signal = 0.03425 + 0.00097 × conc (R² = 0.9996, n = 5); unknowns: 377.1 µg/mL, 841 µg/mL

Check your result: put a standard back through the line: 500 µg/mL gives 0.0343 + 0.000970 × 500 = 0.519, close to the measured 0.53. An unknown’s concentration put back through the line should return its signal: 377 µg/mL gives 0.40.
R² of the opening protein assay
0.9996
R² of the opening protein assay
µg/mL for an absorbance of 0.40
377
µg/mL for an absorbance of 0.40
standards at least
3
standards at least
absorbance of the top standard: higher readings are flagged
1.00
absorbance of the top standard: higher readings are flagged
  • Linear fit with equation and R²
  • Unknown concentrations from signals
  • Flags samples outside the standards
  • Plot of the curve
  • Runs in your browser, free

01

How to use the calculator

  1. Paste the concentrations of your standards, then their signals (absorbance, fluorescence, peak area) in the same order.
  2. Paste the signals of the unknown samples.
  3. Check R² and the plot, then read the concentrations from the table.
  4. Multiply by the dilution factor if the samples were diluted before measuring.
residualxy
A fitted line and a residual. Each dot is one pair of measurements. The line is the one that makes the squared vertical distances as small as possible. A residual, the dashed segment, is the distance from a point to the line; the tighter the points hug the line, the closer r is to 1 or −1.

02

The calculation

signal = intercept + slope × concentration (least-squares fit)
concentration = (signal − intercept) ÷ slope

The line is fitted by ordinary least squares, the same method as a spreadsheet trendline; the linear regression calculator shows the standard errors if you need them.

03

Worked example

A protein assay with five standards:

Standard (µg/mL)01252505001000
Absorbance0.030.160.270.531.00

absorbance = 0.0343 + 0.000970 × concentration, R² = 0.9996

An unknown reading 0.40 is (0.40 − 0.0343) ÷ 0.000970 = 377 µg/mL; one reading 0.85 is 841 µg/mL.

04

Judging the curve

  • R² close to 1 means the standards lie near the line. Many labs look for 0.98 or 0.99 and above; a lower value usually traces to one bad standard or to a curve that bends.
  • Look at the plot. A high R² can hide a systematic bend: points above the line in the middle and below at the ends.
  • Stay inside the standards. The line is only tested between the lowest and highest standard. The calculator marks unknowns beyond them.
  • Blank consistently. Either subtract the blank from every reading or include it as a zero standard.

05

When a straight line is not enough

Many assays are linear only over part of their range. Colorimetric protein assays flatten at high concentration, and ELISA curves are S-shaped and are normally fitted with a four-parameter logistic model, which this calculator does not do. For those, either restrict the standards to the linear part and dilute samples into it, or use software with a 4PL fit. For qPCR, enter log₁₀ of the copy number as the concentration and Ct as the signal: that relationship is linear.

06

Common standard curve mistakes

  • Reading beyond the top standard. An absorbance of 1.30 works out to 1,305 µg/mL, above the 1,000 µg/mL standard; the calculator marks it, and the sample should be diluted and measured again.
  • Forgetting the dilution factor. A sample diluted fourfold that reads 0.40 is 377 µg/mL in the cuvette and 1,508 µg/mL in the original.
  • Forcing the line through zero. With the intercept fixed at 0 the same 0.40 reads 393 µg/mL, 4% higher, because the blank reads 0.03, not 0.
  • Trusting R² alone. The opening curve has R² = 0.9996, yet the 500 µg/mL standard sits 0.011 above the line; look at the plot as well.

Trust

How this page is checked

  • The calculation runs on code with automated tests against worked reference values. Every number in the examples on this page was produced by that same code, not typed by hand.
  • The formulas and conventions used are stated on the page, so you can reproduce the result yourself or compare it with your own software.
  • Last reviewed by the Sciloom team, Zebra Nine Studio Inc. It is a tool for research and teaching, not a substitute for a statistician, a validated lab method or medical advice.

Cite this page

Sciloom. (2026). Standard curve calculator. Zebra Nine Studio Inc. https://sciloom.si/standard-curve-calculator

Read the full methodology · Who makes Sciloom

Frequently asked questions

How do I calculate concentration from a standard curve?

Fit a line to the standards, then for each unknown subtract the intercept from its signal and divide by the slope. The calculator does both steps.

What is a good R² for a standard curve?

Commonly 0.98 or higher, and 0.99 for quantitative work, but also check the plot: a curved pattern can still give a high R².

What if my sample is above the highest standard?

Dilute the sample so its signal falls within the standards, measure again, and multiply the result by the dilution factor.

Should the curve go through zero?

Not forced. A fitted intercept absorbs a small blank offset; forcing the line through zero can bias every result.

Setting up the standards? The serial dilution calculator gives the volumes for each tube.
Free to start, no sign-up.
Back to the tool ↑

More research tools